Fig. 6
From: Efficacy and safety testing of a COVID-19 era emergency ventilator in a healthy rabbit lung model

Representative flow (a) and pressure (b) waveforms generated by the Carestation (green line) and FALCON (orange) sampled over 10 s. VT (c), ΔPIP (d), ΔPEEP (e), RR (f), and I:E ratio (g) derived from the average of 1-min samples taken every 10 min for the duration of ventilation are shown for the Carestation (green) and FALCON (orange). ΔPIP calculated as ΔPIP = PIP − PIPtarget, and ΔPEEP calculated as ΔPEEP = PEEP − PEEPtarget. Either two-tailed paired Students t-test (VT, ΔPIP, ΔPEEP, and I:E ratio) or Wilcoxon signed rank test (RR) were performed (n = 5 rabbits, α = 0.05). ****p < 0.0001, I:E inspiratory time to expiratory time, NS not significant, PEEP positive end expiratory pressure, PIP peak inspiratory pressure, RR respiratory rate, VT tidal volume